ZR-1013 Biosafety Cabinet Tester

Short Description:

ZR-1013 Biosafety cabinet tester adopts potassium iodide (KI) method to test the protection performance of biosafety cabinet class II, conforms to JJF 1815-2020 <Calibration Specification for Class II biosafety Cabinets> and related standards. Build in background test, operator protection, product protection and cross protection four working mode. It is meanly used to test whether the aerosol leaked to outside of the environment, whether the outer agents could enter the work area, whether the contamination among samples is minimized. It is applicable to the protection performance test of class II biosafety cabinets by metrological verification departments, scientific research institutes and biosafety cabinet manufacturers.


Product Detail

Specification

ZR-1013 Biosafety cabinet quality tester adopts potassium iodide (KI) method to test the protection performance of biosafety cabinet class II, conforms to JJF 1815-2020 <Calibration Specification for Class II biosafety Cabinets> and related standards. Build in background test, operator protection, product protection and cross protection four working mode. It is meanly used to test whether the aerosol leaked to outside of the environment, whether the outer agents could enter the work area, whether the contamination among samples is minimized. It is applicable to the protection performance test of class II biosafety cabinets by metrological verification departments, scientific research institutes and biosafety cabinet manufacturers.

Features

> 8-inch colored high definition touch screen, the contents are intuitive and operation is easy.

> Equipped special aerosol sampler, could capture special size aerosol.

> Equipped calibrated special KI aerosol generator, the rotation speed is stable.

> Four path independent high precision sampling module, automatic flow control, don’t need to adjust pressure to stabilize flow.

> Test results could be exported to USB drive or printed by Bluetooth printer.

> High precision liquid supply module, guarantee a stable liquid flow.

> The automatic aerosol control port adopts PID arithmetic, control by real time feedback generator rotation speed.

> Operator protection, product protection and cross protection test three test modes could start after just press the button.

Standards

YY 0569-2011          Class II biological safety cabinets

JJF 1815-2020        Calibration specification for of class II biosafety cabinet

DB52T 1254-2017   Technical practice for testing biosafety cabinets

Deliver Goods

deliver goods Italy
  • Previous:
  • Next:

  • The main parameters

    Parameter range

    Resolution

    Maximum permitted error (MPE)

    Sampling flow

    100L/min

    0.1L/min

    ±2.0%

    Rotation speed of aerosol generator

    28000r/min

    /

    ±500r/min

    Diameter of rotating plate on aerosol generator

    38mm

    /

    /

    Highest position of  X1、Y1

    1000mm

    Sampling membrane

    Diameter 25mm,Aperture 3μm

    Disturbing cylinder

    Diameter 63mm,Length 1100mm

    Noise

    <65dB(A)

    Working temperature

    0-40℃

    Power supply

    AC220V±10%,50Hz

    Overall size

    (length 450×width 380×height 720)mm

    Overall power

    < 400W

    Overall weight

    About 23kg

    Write your message here and send it to us